#1
|
||||
|
||||
|
Quote:
Thanks as well to Bill who confirms that these units do test electrolytics contrary to the statement I quoted from the manual in my prior post. I thought it odd that it is written in the manual and then mine so far isn't testing them. And a complete recap is always a good thing, but I have found not always necessary. There are known problem child caps in these type units like the big 2 mfd non polarized (just to name one) which remains to be seen here. I just haven't had time to hook it up to my Heathkit IT-28 yet. Heck, there's not many caps in the whole darned thing. LOL! Not like I'm doing a 40s RCA TV set......
__________________
"Face piles of trials with smiles, for it riles them to believe that you perceive the web they weave, and keep on thinking free" |
|
#2
|
||||
|
||||
|
The 2mf cap is not in the circuit during leakage testing, so I don't think that it is the cause of of the closed eye problem.
Just a wag, but is the switch on the power factor pot working ok? might try some De-Ox on it. If the switch is closing ok are the 3 resistors connected to it ok and connected correctly? If the 1k was swapped with the 470k, the switch would have little effect on the current supplied to charge the cap in leakage mode. I think that the statement in the manual is just a cya statment, since in some cases a large electrolytic could draw enough current to keep the eye closed (5 to 10 ma) and still function fine... really depends on the circuit. jr |
|
#3
|
||||
|
||||
|
Quote:
![]() I really appreciate the input from both of you guys. I have several cap checkers, but I have never had a Knight and you don't see them much. That statement from the manual held me back from the logical troubleshooting that we are beginning to talk about. Now I know its capability and I should be able to sort this out. As I (for whatever reason) enjoy messing with these old things, I want to develop a better understanding of the circuits and also current flow. Cadillac units like Sprague TelOhmikes' have a panel ammeter for a reason.
__________________
"Face piles of trials with smiles, for it riles them to believe that you perceive the web they weave, and keep on thinking free" |
|
#4
|
||||
|
||||
|
Quote:
With the 10 k resistor, the eye stays closed at every leakage voltage test position when the pf switch is in paper and mica mode, and will open to varying degrees at 150 volts and below in electrolytic mode. With the 10 meg resistor, the eye never closes at any voltage setting when the tester is in electrolytic mode, and will close completely at test voltages above 150 volts in paper and mica mode. The C-3 is about 1000 tomes more sensitive to leakage in the paper and mica mode than it is in electrolytic mode. Your Knight tester is functionally the same as the C-3 and should behave in a similar manner, if the sensitivity is changed correctly by the pf switch. jr |
|
#5
|
||||
|
||||
|
Quote:
"I will check the switch. I assumed it was working as when in bridge mode the pot will alter the eye opening. At first there was flutter as the pot was dirty, but I cleaned that early on. I will have to look at things like the two resistors potentially not being connected correctly. Anything could be connected wrong if the builder installed the wafer function switch upside down. " First, I only verified mechanical opening and closing of the switch. Second, the pot alters the eye in electrolytic mode (I meant that the S2 was in a bridge function at the time). Third, the potentiometer action really has nothing to do with the condition of the switch. The last part of my statement was as it turns out the answer. So now I have a brand new/old Knight cap checker to add to my collection.
__________________
"Face piles of trials with smiles, for it riles them to believe that you perceive the web they weave, and keep on thinking free" |